BITSCOPE product BS120 BITSCOPE MODEL 120 SPECIFICATIONS

The correct appearance of this website is only visible in graphical browsers that support web standards. We try to ensure our content is accessible to any browser but we recommend you upgrade your web browser if you can.
Input Channels Specification
Analog Inputs [1] 4 (2 x BNC + 2 x POD)
Analog Channels [2] 2 (dual A/D converters)
Analog Bandwidth [3] 100 MHz (BNC), 20 MHz (POD)
Logic Channels (POD) [2] 8 (3.3/5V CMOS/TTL)
Input Impedance 1M±1%, 20pF (BNC), 100K/10pF (POD)
Input Coupling AC, DC, GND (BNC), DC (POD)
Input Range (BNC) ±48.8mV ~ ±11.8V in 8 ranges
Input Range (POD) ±462mV ~ ±5.43V in 4 ranges
Input Prescale [4] x20 (BNC only)
Probe Attenuation 1:1 to 3000:1 in 16 ranges
Protocol Capture UART, SPI, I2C & CAN Bus
Vertical Scaling 500 uV/div to 5 V/div (1:1 probe)
Vertical Accuracy ±2% or ±3% (prescale enabled)
Analog Sensitivity 1mV to 40mV (no prescale enabled)
Maximum Sensitivity [5] 300 uV (Time), 80uV (Freq), 20uV (Dithered Mean)
Data Acquisition Specification
Streaming Sample Rates [6] 1Sa/S to 50kSa/S (continuous)
Real-Time Sample Rates [7] 15kSa/S to 40MSa/S (per buffer)
Sub-Sampling Rates [8] 1MSa/S to 1GSa/S (per buffer)
Timebase Range (real-time) 100nS/div ~ 500mS/div
Timebase Range (streaming) 10mS/div ~ 500S/div
Timebase Range (sub-sampling) 1nS/div ~ 1mS/div
Timebase Accuracy 0.01% (100ppm)
Hardware Sample Rates 15kSa/S to 40MSa/S (2000+ rates)
Capture Buffers 64kS x 2 (analog) or 64kS x 9 (analog+logic)
Capture Resolution 8 bit converter with 6.8 to 7.6 ENOB
Maximum Resolution [9] 11 ENOB (at 1MHz and lower)
Display Frame Rate (max) 40 Hz (single channel, 25 mS per analog channel)
Triggers & Timing Specification
Trigger Types Edge (Rise/Fall), Level/State, Logic & Ext.
Trigger Level Range Same as selected Input Range
Hysteresis/Sensitivity ±2% full scale (per range)
Pre-Trigger Hold-Off 10uS ~ 1S
Duration Trigger Filter 10uS ~ 150uS
External Trigger Bus YES (logic inputs may also be used)
Post-Trigger Delay 8 uS to 2 hours (programmable)
Logic Pattern Trigger 8 bit on logic or digitized analog
High Speed Trigger YES (on any analog input)
Glitch Capture (min) 25nS (logic channel or trigger)
Waveform Generation Specification
Generator Types [11] VSR, DDS, CLK & PRNG
Frequency (max) [12] 1MHz (DDS, PRNG), 1.5MHz (VSR) or 5MHz (CLK)
Frequency (min) [12] 1mHz (DDS), 1Hz (VSR, PRNG) or 1kHz (CLK)
Frequency Resolution 6 digits or 1mHz (DDS)
Accuracy [13] ±50ppm, 20° to 30° (typical)
Temperature Coefficient [14]≈1ppm/°C
Built-in Wave-Functions Sine, Ramp, Step, Exponential, Noise
Loadable Wavetable Size 2 to 1024 samples
Generator Sync [15] Specification
Operating Modes Burst, Gated, Sweep, Clock In/Out
Burst Duration [16] 1uS to >1Hr
Post-Trigger Delay [16] 500nS to >1Hr
Inter-burst Delay [16] 500nS to >1Hr
Burst Repeat Count 2^32 (max)
WavePort Specification
Output Level 10Vpp to 40mVpp
Output Offset ±2.5Vpp
Output Impedance 50Ω
Output Voltage Tolerance ±12V (max)
Waveform Resolution (D/A) 8 bits (multiplying)
TimePort Specification
Output Type 3-state, 100k to GND
Logic Levels 5V CMOS In/Out
Output Drive Active (low ≈0.6V, high ≈4.4V)
Output Impedance 100Ω (50mA max)
Trigger Latency 200nS (typical)
External Clock (max) 2MHz
Minimum Pulse Width 100nS
Interfaces Specification
Analog Interface 2 x BNC and 2 x POD
Logic Interface 8 channels I/O (POD)
Control Interface (POD) V+, V-, GND, WavePort, Serial I/O
Front Panel Pins Calibration/TimePort, Trigger Bus, GND
PC Host Interface Opto-Isolated USB 2.0 (USB 1.1 compatible)
Data Upload Speed (max) 1,200 kb/S (Raw Data), 450 kb/S (Enhanced)
Host/General Specification
Included PC Software [10] BitScope DSO Virtual Instrument Software
Optional PC Software [10] Logic, Meter, Chart & Library
Power Requirement 9~12 VDC @ 500mA (max), 2.5mm CP
Operating Temperature 0ºC to 40ºC or 20ºC to 30ºC (for quoted accuracy)
Storage Requirements -40°C to +60°C / 5% to 85% RH
Dimensions (WxDxH)(net) 150 x 35 x 105 mm
Weight (net) 310g
E&OE 
[1] Analog capture for each channel is selectable between the BNC or POD inputs individually.
[2] Dedicated A/D converter for each analog channel to support sample synchronous capture at all supported hardware rates on both analog channels or one analog and all logic channels.
[3] Maximum bandwidth of analog channels captured using equivalent time sampling or used with the multi-band spectrum analyzer via 10:1 probes with prescale disabled and waveform amplitude captured to 10% full-scale.
[4] User selectable analog input prescaler for use with low level signals or when 20MHz input bandwidth limiting is required.
[5] Minimum measurable peak-to-peak sinusoidal waveform voltage when both resolution enhancement and input multiplier are enabled, when viewed as a waveform (Time) or viewed as a spectrum (Freq) and the minimum measurable average DC voltage (Mean).
[6] Streaming capture sample rates for data logging and transient recorder applications.
[7] Real-Time (Nyquist Limited) sample rates (captured per frame).
[8] Equivalent-Time (not Nyquist limited) sample rates sample rates (captured per frame). Applies to periodic signals only.
[9] Specifies the maximum effective number of bits (ENOB) achieveable with resolution enhancement enabled.
[10] All specifications are quoted on the basis that the product is used with the apropriate BitScope software.
[11] Supports 4 different generator types: VSR = Variable Sample Rate, DDS = Direct Digital Synthesis, CLK = Clock Generator, PRNG = Psuedo Random Number Generator.
[12] Minimum and maximum frequency depends on generator type as indicated.
[13] Uncalibrated frequency accuracy is specified (as defined by the BS120U reference crystal). May be calibrated to better than this if precision frequency reference is available.
[14] Temperature Coefficient is that of the BS120U reference crystal.
[15] Trigger, Burst and Clock modes may be configured to accept or generate a trigger or clock signal via the TimePort.
[16] The maximum duration or delay depends on the sample rate. The maximum duration in all cases exceeds 1 hour but can extend far beyond this limit at lower sample rates.

Analog and logic probes are sold seperately.